CH 2 · notes
Assumptions
Each step divides the starting concentration by the dilution factor raised to the step number. Step 1 is the first dilution, not the undiluted stock. The factor is the same at every step, and the final volume of each dilution is the same, so the transfer volume is the same on every row. The page does not draw a plate and it does not vary the factor from row to row.
Concentration inputs are mol/L, mmol/L, µmol/L, and nmol/L. Each step concentration is shown on the mol/L, mmol/L, µmol/L, nmol/L ladder, picking the largest unit whose coefficient is at least 1 and below 1000. Final volume and the minimum pipette volume are litres, millilitres, or microlitres. Before the comparison, both volumes are converted into the final-volume unit. The transfer and the diluent are displayed in that unit unless the number would be scientific notation, in which case a larger or smaller volume unit is chosen. A step is marked only when the transfer is strictly smaller than the minimum. An equal transfer is not marked. The diluent is not marked.
The factor must be greater than 1. The number of steps is an integer of at least 1. Every numeric field must be greater than zero. The first failing field is reported, in the order starting concentration, dilution factor, number of steps, final volume, minimum pipette volume.
The diluent is not an input. The transfer volume is the volume of the previous concentration you move into the next tube. Diluent volume is the final volume minus that transfer, and it is the same on every row. You then bring that tube to the final volume you typed. The page does not subtract the aliquot a second time. The final volume is the volume of the finished dilution, including the transferred volume. Concentrations, transfer, and diluent use 4 significant figures, with ties rounded half away from zero. On-screen numbers use a plain number or a power of ten. That display rule is not a statement about pipette accuracy.
How it works
Cn = C0 ÷ Fn
Vtransfer = Vfinal ÷ F
Vdiluent = Vfinal − Vtransfer
n = 1…N
The plate engraves those three relations. The working panel, once the inputs are valid, writes one concentration line per step, one transfer line, and one diluent line. Each concentration string, the transfer string, and the diluent string are the strings already returned for the step table. The tape does not recompute them with a different rounding rule. Copy writes the step table as tab-separated rows, with the last column yes or no. That clipboard text is separate from the on-screen badge.
Changing any input recalculates immediately. An invalid factor clears the table. The copy button disables with it and will not write the previous table.
Worked example
Starting concentration 8 µmol/L. Dilution factor 4. Steps 3. Final volume 200 µL. Minimum pipette volume 60 µL.
- Step 1: 2.000 µmol/L. Transfer 50.00 µL. Diluent 150.0 µL. Below minimum pipette volume: yes.
- Step 2: 500.0 nmol/L. Transfer 50.00 µL. Diluent 150.0 µL. Below minimum pipette volume: yes.
- Step 3: 125.0 nmol/L. Transfer 50.00 µL. Diluent 150.0 µL. Below minimum pipette volume: yes.
- C1 = 8 µmol/L ÷ 4^1 = 2.000 µmol/L
- C2 = 8 µmol/L ÷ 4^2 = 500.0 nmol/L
- C3 = 8 µmol/L ÷ 4^3 = 125.0 nmol/L
- Vtransfer = 200 µL ÷ 4 = 50.00 µL
- Vdiluent = 200 µL − 50.00 µL = 150.0 µL
Because the factor and the final volume do not change with the step, every row carries the same transfer. In this example that transfer is under the minimum, so every row is marked. Raise the final volume or lower the factor if you need a transfer your pipette can make. The concentrations above are the function's step strings.
Common mistakes
- Counting the stock as step 1. Step 1 is already diluted by one factor.
- Expecting a different transfer or diluent on later rows. Only the concentration changes.
- Treating an equal transfer as too small. The mark is strict: less than the minimum, not less than or equal.
- Comparing a millilitre minimum with a microlitre final volume without letting the page convert them. Enter each volume with its own unit. The comparison uses the final-volume unit.
- Using this page to find the mass of a stock. That is the molarity channel. This page starts from a concentration you already have.
FAQ
Why is the transfer volume the same on every row?
Each row uses the same dilution factor and the same final volume. Transfer volume is final volume divided by that factor, and diluent volume is final volume minus that transfer, so neither changes with the step number. Only the concentration changes.
When is a step marked below the pipette minimum?
Only when the transfer volume is strictly less than the minimum you entered. An equal transfer is not marked. The diluent volume is not compared with the pipette minimum. Both volumes are converted into the final-volume unit before that comparison.
Is this a plate map?
No. The output is a step list: concentration, transfer volume, and diluent volume for each step. It does not draw wells.
What if the dilution factor is 1?
A factor of 1 or less is rejected. The field message is Enter a valid dilution factor. The previous table, if you had one, is cleared.
Other channels
The stock concentration comes from Molarity when you are weighing a solid.Primer Tm does not use this step list. All three channels are indexed on the home page.