Tm (°C) = 1000 ΔH / (ΔS + 1.987 ln C) − 273.15

SantaLucia 1998, 250 nM, 50 mM Na+ default. Mg2+ uses Owczarzy 2008.

Method notes

ΔS salt term = 0.368 × (N − 1) × ln[Na+]

Na+ is mol/L (entered mM × 10⁻³). Blank salt uses 50 mM.

C = 250 × 10⁻⁹ mol/L when primer concentration is blank. Complement neglected.

ΔH is kcal/mol; ×1000 converts it to cal/mol.

nearest-neighbor

SantaLucia 1998 unified parameters. No symmetry term.

Blank Mg2+ keeps saltcorr 5. A filled Mg2+ uses Owczarzy 2008, saltcorr 7, unless dNTP is at least that Mg2+.

Suggested annealing is the lower Tm minus 5 °C. It is a starting point, not a kit protocol.

Calculation values stay on this device.

CH 4 · notes

Assumptions

The channel returns one melting temperature, in degrees Celsius, to one decimal place. The sequence is DNA bases A, C, G, and T, case-insensitive. Anything else is rejected before the length check. A sequence that is only those bases and is shorter than 10 bases returns no temperature. Salt is optional and is a monovalent cation concentration in millimolar. Leave it blank and the same equation uses 50 mM Na+. A filled salt must be greater than zero.

Primer concentration is optional. Leave it blank and the equation uses 250 nM. The complementary strand is ignored. That is the excess-primer case, not the equal-strand case. Mg2+ and dNTP are optional. Leave Mg2+ blank and there is no magnesium term. A reverse primer is optional. The suggested annealing temperature is the lower Tm minus 5 °C. It is a starting point, not a polymerase-kit setpoint, and it is not a clinical or dosing figure. If the entered dNTP is at least the entered Mg2+, the monovalent correction is used and a note says so. If the two Tms differ by more than 5 °C, a warning asks you to consider redesigning one primer. The annealing temperature is still the lower Tm minus 5 °C.

How it works

Tm (°C) = 1000 ΔH / (ΔS + 1.987 ln C) − 273.15

SantaLucia 1998, 250 nM, 50 mM Na+ default. Mg2+ uses Owczarzy 2008.

ΔS salt term = 0.368 × (N − 1) × ln[Na+]
Na+ is mol/L (entered mM × 10⁻³). Blank salt uses 50 mM.
C = 250 × 10⁻⁹ mol/L when primer concentration is blank. Complement neglected.
ΔH is kcal/mol; ×1000 converts it to cal/mol.
nearest-neighbor
SantaLucia 1998 unified parameters. No symmetry term.
Blank Mg2+ keeps saltcorr 5. A filled Mg2+ uses Owczarzy 2008, saltcorr 7, unless dNTP is at least that Mg2+.
Suggested annealing is the lower Tm minus 5 °C. It is a starting point, not a kit protocol.

The plate above carries the same equation and the same parameter line. The notes under the working panel are the same sentences. With Mg2+ left blank, salt correction is the SantaLucia 1998 entropy term, 0.368 × (N − 1) × ln[Na+], which is saltcorr 5 in the BioPython naming used to check this function. It is not the Owczarzy 2004 correction. A filled Mg2+ switches to Owczarzy 2008, saltcorr 7, unless the entered dNTP is at least that Mg2+. The magnesium path is not part of the comparison below. There is no symmetry term, because self-complementarity is off. ΔH from the parameter table is in kcal/mol, and the factor 1000 converts it to cal/mol before dividing by the gas constant 1.987 cal/(mol·K).

Two conventions move the number. Both are recomputed for this primer at 50 mM Na+, with no Mg2+, at the default 250 nM. The concentration term on this channel is C: the primer is in large excess and the complement is neglected. The equal-strand convention uses C/4 in the same logarithm. That change alone shifts this primer by about −1.6 °C. The salt term on this channel is SantaLucia 1998. IDT OligoAnalyzer documents the Owczarzy 2004 correction, and the 2008 correction when Mg2+ is present. Substituting Owczarzy 2004, still at C, shifts this primer by about +1.0 °C.

The two effects partly cancel. C/4 together with Owczarzy 2004 leaves this primer about −0.7 °C from the Tm on this page. That pair is not an IDT result and not an NEB result. NEB's Tm API states that its values use SantaLucia 1998 data with the Owczarzy 2004 salt correction, and Schildkraut's correction for Phusion. Primer concentration on that API is a required input in mM. Published examples pass 0.5, 0.4, or 0.2, depending on the request. Each tool uses different default parameters. Match their settings before comparing.

Worked example

pUC/M13 forward 24-mer, sequence CGCCAGGGTTTTCCCAGTCACGAC, length 24. GC 62.5%. Salt is left blank, so the function uses its blank-salt path.

  1. method = nearest-neighbor, SantaLucia 1998, saltcorr=5
  2. salt = 50 mM Na+
  3. primer = 250 nM, complement ignored
  4. Tm = 65.4 °C

Displayed Tm: 65.4 °C. That string is the function return for this sequence and a blank salt. It is not a temperature copied from another calculator, and it is not an annealing setpoint.

Common mistakes

  • Pasting a sequence that still contains spaces, numbers, or U. Those characters fail the alphabet check before length is considered.
  • Reading a blank salt field as zero. Blank means 50 mM Na+ in the same equation.
  • Comparing the display with an NEB or IDT result that used Owczarzy salt correction, a different primer concentration, or added Mg2+ and dNTPs.
  • Programming the thermocycler to this Tm without choosing an annealing offset for the polymerase and buffer you actually run.
  • Expecting GC% or the base count on the working panel. Both are on the LCD, next to the Tm. With the default settings the tape still repeats the method name, the salt, the 250 nM primer, and this Tm. Suggested annealing is on the LCD. It is also on the tape when a reverse primer is entered.

FAQ

Why can this Tm differ from NEB or IDT?

Two conventions, recomputed for the worked example at 50 mM Na+ with no Mg2+. This channel uses C, primer in large excess. The equal-strand convention uses C/4 and shifts this primer by about −1.6 °C. This channel uses SantaLucia 1998 salt correction on that default. IDT OligoAnalyzer documents Owczarzy 2004, and Owczarzy 2008 when Mg2+ is present. Owczarzy 2004 alone shifts this primer by about +1.0 °C. The two effects partly cancel, net about −0.7 °C. Those shifts are not an IDT or NEB output. A filled Mg2+ on this page uses Owczarzy 2008, unless dNTP is at least that Mg2+, and is not part of that comparison. NEB's Tm API uses SantaLucia 1998 data with Owczarzy 2004, and Schildkraut for Phusion. Each tool uses different default parameters. Match their settings before comparing.

What does a blank salt field mean?

Blank monovalent salt is not zero. The same nearest-neighbor equation runs with 50 mM Na+. A filled value must be a number greater than zero. It is the total of Na+ and K+.

Why is there no Tm for a very short oligo?

A sequence of only A, C, G, and T that is shorter than 10 bases returns no Tm. The message is Sequence too short for a reliable Tm (minimum 10 bases). Any other character is rejected first, as an invalid DNA sequence.

Is this Tm the annealing temperature I should program?

The large number is the nearest-neighbor Tm. The page also shows a suggested annealing temperature: that Tm minus 5 °C, or the lower of the two Tms minus 5 °C when a reverse primer is entered. If those Tms differ by more than 5 °C, the page says so. The annealing figure is still the lower Tm minus 5 °C. It is a starting point for a gradient, not a polymerase-kit setpoint.