Molarity

Solve for concentration, mass or volume, or convert g/mL or normality to molarity.

live · runs in your browser

INPUT 3 / 3 set

Before you weigh

  • Use the molecular weight on the CoA or label for the exact form you weigh (salt, hydrate, free base). For example, an HCl salt or a dihydrate changes the MW; tris base 121.14 g/mol is not the same solid as tris HCl.
  • Purity or assay correction: mass to weigh = target mass ÷ (purity as a fraction). Check whether the CoA assay is on an as-is or anhydrous/dried basis. If the assay is on the anhydrous/dried basis, convert first: as-is purity = assay × (1 − water or LOD fraction). For example, 98.0% as-is for a 5.844 g target means weigh 5.963 g.
  • For normality, n depends on the specific reaction (acid–base or redox) and is not a fixed property of the compound. For example, H₂SO₄ may be n = 2 in full neutralization and n = 1 in a one-proton step.

CONC · C

Enter mass, molecular weight and volume to see the concentration.

c = m ÷ (MW × V)

c (mol/L) = mass_g / (MW_g_per_mol × volume_L)

Calculation values stay on this device.

CH 1 · notes

Assumptions

This channel solves one equation. You choose the unknown with Solve for. The default is concentration, from a weighed mass, a molecular weight you type, and the final volume. Mass and volume use the same equation rearranged. It does not build a dilution series. Activity coefficients, density, and temperature are outside the equation, so a concentration is the formal concentration of what you weighed, not a thermodynamic activity and not a clinical dose. A calculated mass is the mass that equation asks you to weigh. It is not a clinical dose.

Mass is grams, milligrams, micrograms, or nanograms. Volume is litres, millilitres, or microlitres. Concentration is mol/L, mmol/L, µmol/L, or nmol/L. A smaller unit is divided by its scale before the equation: 1000 for milli, 1000000 for micro, and 1000000000 for nano. mmol/L is the mol/L figure times 1000. The display keeps 4 significant figures, with ties rounded half away from zero. A blank, a non-number, or a value that is not greater than zero stops the channel and names the field. The first failing field depends on the unknown. Concentration checks mass, then molecular weight, then volume. Mass checks concentration, then molecular weight, then volume. Volume checks mass, then molecular weight, then concentration.

Mass concentration (g/L, mg/mL, µg/mL, or % w/v) converts in the Mass concentration ↔ molarity mode. Oligo concentration from a delivered nmol or OD260 amount uses a different equation.

The molecular weight is an input. The worked example uses anhydrous sodium chloride. A hydrate is a different compound. If the bottle is a hydrate, look up that hydrate on PubChem and type its molecular weight. Do not reuse the anhydrous figure.

How it works

c = m ÷ (MW × V)

c (mol/L) = mass_g / (MW_g_per_mol × volume_L)

Mass is in grams and volume is in litres inside that equation. The plate above engraves the same relation. After a valid entry, the working panel writes the substitution with strings from this same function: the volume conversion when the volume was entered in millilitres or microlitres, the amount of substance as mass divided by molecular weight, and concentration as that amount divided by the volume in litres. The other concentration unit is a second call with the unit switched. It is shown beside the main result and is not copied to the clipboard.

Nothing in the tape is a second formula. If you change a unit or the unknown, the page recalculates immediately. There is no submit button. Copy result writes the displayed result and its unit for the quantity you are solving for. In the default mode that string is the concentration. Copy protocol writes the prep steps and the working lines for that same mode. It does not replace the Copy result string. Print protocol uses the page's print stylesheet.

Worked example

anhydrous sodium chloride, NaCl, molecular weight 58.44 g/mol (PubChem CID 5234). Mass 2.922 g. Final volume 100 mL.

  1. V = 100 mL ÷ 1000 mL/L = 0.1000 L
  2. n = 2.922 g ÷ 58.44 g/mol = 0.05000 mol
  3. c = 0.05000 mol ÷ 0.1000 L = 0.5000 mol/L

Displayed concentration: 0.5000 mol/L. The other unit, from the second call, is 500.0 mmol/L. Both strings are the function's display, not a rounded figure typed beside it. The volume in the tape is the final solution volume. It is not the volume of water you poured in before the salt dissolved. This example solves for concentration, which is the default. The same equation also solves for mass or for volume when you change Solve for.

Common mistakes

  • Typing the anhydrous molecular weight while weighing a hydrate, or the reverse. The PubChem record has to match the solid.
  • Entering the water you added instead of the final volume in the flask. The same bring-to-the-mark step is how to make 70% ethanol.
  • Leaving the mass unit on milligrams while reading the number as grams. The unit control is part of the input. The tape writes the conversion when the unit is smaller than grams or litres.
  • Treating 4 significant figures as a statement about balance tolerance or reagent purity. The figures are the display rule.
  • Planning the dilution steps on this page. The step list and the pipette-minimum mark are on the serial dilution channel.
  • Leaving Solve for on concentration when the number you want is the mass. The default answers with a concentration. Switch to Mass before reading the display as grams.

FAQ

How do I convert g/mL to molarity on this page?

Choose Mode Mass concentration ↔ molarity. Enter the mass concentration, its unit, and the molecular weight in g/mol. The page converts to g/L first (% w/v is grams per 100 mL, so 1 % w/v = 10 g/L), then divides by MW to get mol/L. You can also solve for mass concentration when molarity and MW are known. Display uses the same mol/L to nmol/L ladder as the concentration solver.

How does normality relate to molarity here?

Choose Mode Normality ↔ molarity. Enter normality N in eq/L and n, the equivalents per formula unit (a positive number). Molarity is M = N ÷ n; normality is N = M × n. n depends on the reaction—for example H2SO4 is often 2 in full acid–base neutralization and HCl is 1. The page does not pick n for you.

Can this page tell me how much to weigh?

Yes, when Solve for is Mass. Enter the concentration, the molecular weight, and the final volume. The mass is that equation rearranged, at 4 significant figures. Prep steps weigh that calculated mass. They do not use a second formula. The default is still concentration. In that mode the prep steps repeat the mass you typed.

What molecular weight should I type for sodium chloride?

Type the molecular weight of the solid in the bottle. The worked example uses anhydrous NaCl, 58.44 g/mol, PubChem CID 5234. A hydrate is a different solid. Use that hydrate's own PubChem record, and do not reuse 58.44.

Why does mmol/L not look like my mol/L number with three zeros?

mmol/L is the mol/L result times 1000, then shown with the same 4-significant-figure rule. The second unit is another call to the same function, not a hand-edited string.

Does the page upload the numbers?

No. The calculation stays in the browser. The panel notice is the data statement for this page.

Sources

Other channels

With a stock concentration in hand, Serial Dilution builds the step list and compares the transfer with your pipette minimum. Primer Tm is a separate channel. It does not use this concentration. The index of all three channels is on the home page.